Recall Class: Class I
Date Recall Initiated: October 28, 2013
Products: In-Line Dual Stage Ultra (DSU) Filter
These affected products (Item #s 70-0230, 70-0231, 70-0232, 70-0234 and 70-0235) were distributed from January, 2011 through October, 2013.
Use: The Nephros In-Line DSU Filter is a disposable filter that delivers microbiologically pure water for washing and drinking. The DSU Filter is not a complete water treatment system but it serves to remove biological contaminants.
Recalling Firm:
Nephros Inc.
41 Grand Ave.
River Edge, NJ 07661-1947
Manufacturer:
Nephros Inc.
41 Grand Ave.
River Edge, NJ 07661-1947
Reason for Recall: Promotional materials for non-medical water filtration products were determined to promote claims which constitute marketing the product as a medical device. The affected marketing materials include the following document numbers:
- 60-0237 Nephros Ultrafilters Technical Dossier (Infection Control)
- 60-3003 Nephros In-Line DSU Filter Brochure
Public Contact: For questions about this action you may contact Nephros Inc. at: 1-201-343-5202 x100 between the hours of 9:30 a.m. to 5 p.m. ET, Monday through Friday.
FDA District: New Jersey District Office
FDA Comments:
Health Care Professionals: No action is required beyond the recommendations provided in the Urgent Medical Device Recall Notification ![]()
On October 28, 2013, Nephros sent customers recall letters and return response forms via UPS. Customers are asked to provide a contact information list and remove and return filters. All related promotional materials should be discarded.
For questions about this action you may contact Nephros Inc. at: 1-201-343-5202 x100 between the hours of 9:30 a.m. to 5 p.m. ET, Monday through Friday.
Health care professionals and consumers may report adverse reactions or quality problems they experienced using these products to MedWatch: The FDA Safety Information and Adverse Event Reporting Program either online, by regular mail, or by FAX.