Document issued on: March 28, 2013.
The draft of this document was issued on January 5, 2011.
For questions regarding this document, contact Prasad Rao, Ph.D. at 301-796-6203 or [email protected]

U.S. Department of Health and Human Services
Food and Drug Administration
Center for Devices and Radiological Health
Office of In Vitro Diagnostics and Radiological Health
Division of Microbiology Devices
Preface
Public Comment
You may submit written comments and suggestions at any time for Agency consideration to the Division of Dockets Management, Food and Drug Administration, 5630 Fishers Lane, rm. 1061, (HFA-305), Rockville, MD, 20852. Submit electronic comments to http://www.regulations.gov. Identify all comments with the docket number listed in the notice of availability that publishes in the Federal Register. Comments may not be acted upon by the Agency until the document is next revised or updated.
Additional Copies
Additional copies are available from the Internet. You may also send an e-mail request to [email protected] to receive an electronic copy of the guidance or send a fax request to 301-827-8149 to receive a hard copy. Please use the document number 1721 to identify the guidance you are requesting.
Table of Contents
- I. Introduction
- II. Background
- III. Scope
- IV. Risks To Health
- V. Establishing Performance Characteristics
- A. Controls
- B. Performance Studies
- C. CLIA Waiver
- VI. References
Establishing the Performance Characteristics of In Vitro Diagnostic Devices for the Detection of Antibodies to Borrelia burgdorferi
Guidance for Industry and Food and Drug Administration Staff
Premarket Notification 510(k).” Guidance on the content and format for abbreviated and traditional 510(k)s is available in the guidance entitled “Format for Traditional and Abbreviated 510(k)s.”
Information regarding the use of standards can be found in section 514(c)(1)(B) of the FDC Act (21 U.S.C. 360d(c)(1)(B)), and in the FDA guidance entitled “Use of Standards in Substantial Equivalence Determinations,” The Special 510(k) is an option for manufacturers considering modifications to their own cleared devices. Information on how to prepare a Special 510(k) is available at “How To Prepare A Special 510(k).”
Further information on device testing can be found in the guidance entitled “In Vitro Diagnostic (IVD) Device Studies – Frequently Asked Questions” and the guidance entitled “Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable.”
1 devices, you include a larger number of devices in your evaluation, in order to best represent the settings in which the devices will be used.
CLSI document, EP15-A2 [Ref. 6], contains additional information on reproducibility study design.
3. Specimen Collection and Handling Conditions
We recommend that you substantiate statements in your labeling about specimen storage and transport by assessing whether the device can maintain acceptable performance over the range of storage times and temperatures recommended to users. For example, an appropriate study would include an analysis of aliquots stored under the recommended conditions of time, temperature, or specified number of freeze/thaw cycles. We recommend that you state the criteria for an acceptable range of recoveries under the recommended storage and handling conditions as described in the CLSI document, H18-A [Ref. 7].
4. Assay Cut-Off Point
We suggest that data be furnished to explain how the assay cut-off point was selected and established. If appropriate, information should be provided on the use of an equivocal zone for testing. If data suggests that an equivocal zone is not appropriate, this should be explained.
5. Clinical Performance Studies
Current recommendation for Lyme disease serology testing is a two step procedure [Ref. 1, 2]:
Step I: EIA tests for IgM and/or IgG antibodies
Step II: Western blot tests for IgM or IgG antibodies when EIA tests are positive or equivocal
Since the Borrelia burgdorferi strains from the United States (U.S.) differ markedly from Europe and other non-U.S. endemic regions, all antigenic reagents used in Lyme disease devices to be marketed in the U.S. should be manufactured from the U.S. strain sensu stricto B31 Borrelia burgdorferi spirochete and not from other non-U.S. strains. This is particularly true when whole cell sonicate is used as the antigenic reagent to capture B. burgdorferi antibodies in both EIA and Western blot devices. However, it is acceptable to augment antigens from the U.S. strain of B. burgdorferi with recombinant proteins of defined length but not total cell lysates from the European species of Borrelia. This applies to first tier EIA tests only and not to Western blot tests for IgM or IgG antibodies.
Clinical studies are necessary to establish the performance characteristics for both EIA and Western blot devices. Generally, we recommend that performance be assessed in the testing environment in which the device will ultimately be used (i.e., clinical laboratory) by individuals who will use the test in clinical practice (e.g., trained technologists). The following sets forth what is expected in terms of establishing the performance characteristics for each type of device.
Study Protocol
We recommend that you develop a detailed study protocol that includes specific patient inclusion and exclusion criteria, the type and number of specimens needed, directions for use, and a statistical analysis plan that accounts for variances to prevent data bias. You need to provide a statistical justification to support the sample size of the study population. We recommend that you include this and any other relevant protocol information in your premarket submission.
We encourage sponsors to contact the Division of Microbiology Devices to request a review of their proposed studies and selection of specimen types prior to study initiation.
Study Sites
We recommend that you conduct your studies at a minimum of three separate sites, one of which may be in-house. Note, we do not believe non-U.S. clinical testing site data are appropriate given that different strains are found in other countries. Clinical investigations of unapproved and uncleared in vitro diagnostic devices, including diagnostic devices for Lyme disease, are subject to the investigational device exemption (IDE) provisions of Section 520(g) of the Federal Food, Drug, and Cosmetic Act (21 U.S.C. 360j) and the implementing regulations. You should consider how 21 CFR part 812 (IDEs) applies to your particular study and refer to 21 CFR part 50 (informed consent), and 21 CFR part 56 (institutional review board review) for other applicable requirements.
We recommend that the performance evaluation for devices intended for point-of-care (POC) use include, at a minimum, one site at a clinical laboratory as well as 2 or more sites representative of non-laboratory settings in which the device is intended to be used (e.g., an emergency department).
Study Design
The following studies should be conducted and submitted with a 510(k) application for either a Lyme disease EIA device or a Lyme disease Western blot device. Study designs are similar for both IgM and IgG tests.
a. Sensitivity Study: A study consisting of a minimum of 100 well characterized clinically or culture confirmed Lyme disease specimens should be conducted with the test device. These archived specimens should contain samples from early, early disseminated, and late phases of the disease. The sensitivity of the test device on these specimens should be compared to a predicate device. A pedigree for each characterized patient contributing samples for the studies should be included; this pedigree should encompass available clinical and laboratory testing information.
The following breakout of the patient population is suggested for sensitivity testing:
a. Initial (acute) samples from patients with documented erythema migrans (EM) or culture positive disease (1 month, 1-2 months, 2–3 months after symptom onset).
b. Convalescent samples from patients with documented EM or culture positive results, stratified by time of draw after initial appearance of symptoms (3-12 months).
c. Testing of known Lyme disease patients with presentations other than EM, e.g., neuro-, arthritic, etc., in which the time interval from infection to symptom onset may be unknown or more than 1 year.
Testing of specimens from patients across all ranges is recommended.
b. Prospective Study: A study designed to test non-selected, prospectively collected consecutive specimens from a minimum of three geographically distinct locations within the U.S. should be conducted. Testing of these samples using the test device and a comparison device, should occur at a minimum of 3 sites, one of which could be internal.
i. For EIA devices: The results should be compared to a predicate device. All positive and equivocal samples by the test device and the predicate device should be tested by an FDA-cleared IgM and/or IgG Western blot assay as applicable.
ii. For Western blot devices: The prospective study samples should be tested initially by an FDA cleared first-step EIA. All EIA positive and equivocal samples should be tested by the test device and the predicate Western blot. Positive and negative % agreement between the two devices should be provided.
Interpretation of Western blot results: The interpretation of Western blot results should follow the recommended criteria described by the Centers for Disease Control (CDC) [Ref. 1] and the Second National Conference on Serological Diagnosis of Lyme Disease [Ref. 2]. An IgM immunoblot is considered positive if two of the following three bands are present: 23 kDa (OspC), 39 kDa (BmpA), and 41 kDa (Fla). An IgG immunoblot is considered positive if five of the following 10 bands are present: 18 kDa, 23 kDa (OspC), 28 kDa, 30 kDa, 39 kDa (BmpA), 41 kDa (Fla), 45 kDa, 58 kDa, 66 kDa, and 93 kDa.
Analytical Specificity Testing: For the determination of analytical specificity, testing of samples from both endemic and non-endemic regions is recommended (minimum of 100 samples from each region). These samples should be obtained from an asymptomatic population but should not include pre-screened blood donors. The results, expressed as % positives and % negatives, should be presented separately as endemic and non-endemic subjects.
CDC Reference Panel Testing: A standard panel of positive and negative specimens provided by the CDC for testing Lyme disease detection devices should be tested by the new device and the results provided as % agreement with the expected results. The data should be stratified by disease stage; early, intermediate, and late and be compared to the predicate device.
6. Labeling
Proposed labeling for the device must comply with the requirements of 21 CFR 809.10. As part of meeting these requirements we recommend that your labeling include a description of quality control recommendations, which should include a clear explanation of what controls and calibrators are to be used with the assay and how often they should be used.
2 we recommend that you consult with the Division of Microbiology Devices staff regarding the design of specific studies to support the CLIA waiver application for your device and review “Guidance for Industry and FDA Staff, Recommendations: Clinical Laboratory Improvement Amendments of 1988 (CLIA) Waiver Applications for Manufacturers of In Vitro Diagnostic Devices.”
VI. References
- CDC. Recommendations for test performance and interpretation from the Second National Conference on Serologic Diagnosis of Lyme Disease. MMWR 1995;44:590-591.
- Association of State and Territorial Public Health Laboratory Directors and the Centers for Disease Control and Prevention. Recommendations. In: Proceedings of the Second National Conference on Serologic Diagnosis of Lyme Disease (Dearborn, Michigan). Washington, DC: Association of State and Territorial Public Health Laboratory Directors 1995; 1-5.
- Clinical and Laboratory Standards Institute. 2005. Interference Testing in Clinical Chemistry; Approved Guideline. EP7-A2. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 2004. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline. EP5-A2. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 2002. User Protocol for Evaluation of Qualitative Test Performance; Approved Guideline. EP12-A. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 2006. User Verification of Performance for Precision and Trueness; Approved Guideline. EP15-A2. Clinical and Laboratory Standards Institute, Wayne PA.
- Clinical and Laboratory Standards Institute. 1990. Procedures for Handling and Processing of Blood Specimens; Approved Guideline. H18-A. Clinical and Laboratory Standards Institute, Wayne PA.
1Point-of-care tests, also known as bedside or near-patient tests, is a term that encompasses any diagnostic testing near the site of patient care regardless of whether the device is intended for use by a trained medical professional or by a lay user. The person conducting the diagnostic testing near the point of care, whether a trained medical professional or a lay user, is the caregiver in that instance as that person is the person providing care.
Â